Trilogy200

Ventilator

Trilogy200's advanced flow triggering is ideal for patients who require enhanced inspiratory triggering and leak compensation.

Specifications Show all

Environmental
Temperature
5 – 40 °C
Relative humidity
15 - 95 %
Atmospheric pressure
110 to 60 kPa
Controls
Circuit types
Active with PAP; passive; active with flow
Ventilation types
Volume control; pressure control; bilevel
Volume modes
AC; SIMV (w/PS); CV
Pressure modes
CPAP; S; S/T; T; PC-SIMV (w/PS)
IPAP
4 to 50 cm H₂O
EPAP for active circuits
0 to 25 cm H₂O
EPAP for passive circuits
4 to 50 cm H₂O
CPAP for passive circuits
4 to 20 cm H₂O
PEEP for active circuits
0 to 25 cm H₂O
PEEP for passive circuits
4 to 25 cm H₂O
Pressure support differential
0 to 30 cm H₂O
Physical
Weight
5 kg (11 lb)
Size
16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.5" H)
Alarms
Apnea rate
Off; 10 to 60 sec
Circuit disconnect
Off; 5 to 60 sec
High tidal volume
Off; 50 to 2000 ml
Low tidal volume
Off; 40 to 2000 ml
High minute ventilation
Off; 0.5 to 99 l/min
Low minute ventilations
Off; 0.1 to 99 l/min
High respiratory rate
Off; 4 to 80 beats per minute
Low respiratory rate
Off; 4 to 80 beats per minute
Control accuracy
Breath rate
1 to 60 for AC mode; 1 to 60 for all other modes
Flow trigger sensitivity
1 to 9 l/min
Tidal volume
50 to 2000 ml
Ramp start pressure
0 to 25 cm H2O for active circuits; 4 to 25 cm H2O for passive circuits; 4 to 19 cm H2O in CPAP cm H₂O
Inspiratory time
0.3 to 5.0 sec
Rise time
1 to 6
Flex
Off; 1 to 3
Flow cycle
10 to 90 %
Ramp length
Off; 5 to 45 min
Apnea rate
4 to 60 beats per minute
Measured patient parameters
Respiratory rate
0 to 80 beats per minute
Tidal volume
0 to 2000 ml
Minute ventilation
0 to 99 l/min
Leak rate
0 to 2000 l/min
Peak inspiratory flow
0 to 2000 l/min
Peak inspiratory pressure
0 to 99 cm H₂O
Mean airway pressure
0 to 99 cm H₂O
Percentage patient triggered breaths
0 to 100 %
I:E ratio
9.9:1 to 1:9.9